NeuroTrac™ MyoPlus
|
       
|
|
| 
|
Quick
Links:
Clinical purpose of MyoPlus?
Applications
How does EMG help?
How does ETS help?
How does STIM help?
How to Purchase
NeuroTrac™
Software
MyoPlus
Specifications: |
| Number of channels: |
1 STIM, 1 EMG |
| Number of programms: |
67 |
| Number of phases: |
upto 5 |
| Programme time: |
upto 8 hours |
| Frequency: |
2 - 100 Hz |
| Pulse Width: |
50 - 450 uS |
| STIM Work/Rest: |
from 0.5 sec |
| Stimulation: |
upto 90 mA |
| EMG precision: |
0.1 uV |
|
Features and
Benefits
Portable Desk Top device.
Combined single channel EMG/ Muscle Stimulation. Supplied
with a desk top stand.
EMG + Stimulation mode {Passive + Active}, EMG Triggered
Stimulation, Muscle Stimulation and EMG Modes.
Designed to be used for a wide
range of Physiotherapy, Sports and Rehabilitation
applications including Continence and Stroke treatment.
Custom programs enabling up
to 5 phases to be set up in one program.
Optional Multilingual PC Database
Software
Used as a learning tool with emphasis on improving treatment
protocols using EMG and Neuromuscular Stimulation.
Locks programmes and records Patient
Home Compliance on a daily basis.
Portable, light and user friendly.
|
|
Clinical
Purpose of MyoPlus
EMG Biofeedback
and Neuromuscular Stimulation, separately or in combination
are increasingly being used more to treat
neuromuscular re-education in a wide range of Physiotherapy,
Sports and Rehabilitation applications. The Combination
of EMG plus Neuromuscular Stimulation is now being used
successfully to treat a growing number of maladies such
as Incontinence, in
particular Stress, Prolapses, Urge, lack of sensitivity
and weak Pelvic muscles; Stroke,
Sports injuries, Patellofemoral Pain, Bells Palsy, Back
pain and a wide range of muscular problems.
|
|
|
Applications
Urinary Incontinence
Rectal Incontinence
Pelvic floor pain
Sports Injuries
Migraine Headaches
Chronic tension headaches |
Patella femoral pain (Aqua & Land)
Phantom limb pain
Post-traumatic stress disorders
Neck and shoulder girdle pain
Repetitive strain injury
Asthma
Lower back pain |
Muscle monitoring
Stroke
Neuro feedback |
|
|
EMG
Biofeedback
The Myo Plus can be used as a stand -
alone Biofeedback device using the LCD screen to set up
the appropriate parameters for each individual user or used
in conjunction with the established multilingual software.
The Myo Plus can be set up to:-
Produce structured patient individual sessions or progress
reports monitored on a daily or weekly basis. This helps
the Clinician to analyze and evaluate ongoing treatment;
Enables clinician to record patient work/rest sessions to
verify home compliance and used to reinforce the overall
patient progress;
Used as an assessment tool or purely as a biofeedback device;
Work in conjunction with a computer or as a portable device.
|
|
ETS
EMG -Triggered Stimulation
EMG-Triggered Stimulation [ETS] is a combination
of EMG and Stimulation where the EMG
is used as a threshold to trigger the stimulation. The
patient must achieve a pre-set EMG threshold before the
neuromuscular stimulation phase is triggered. At the end
of each work phase the device reverts back to the EMG rest
phase, stimulation cannot be re-triggered until the patient
achieves the pre-set EMG threshold level. In the automatic
mode the device will continually monitor the work efforts
of the patient. If the patient finds achieving the threshold
to difficult the device will automatically reduce the threshold,
conversely if the patient finds achieving the threshold
to easy, the device will increase the threshold level accordingly.
The Myo Plus can be used as a stand - alone ETS device using
the LCD screen or used in conjunction with the established
Multilingual software.
|
|
STIM
-Neuromuscular Stimulation
It is important that the correct parameters
are set when using stimulation, the EMG diagnosis helps
to determine which are the most effective modes to set or
pre-set program to select. The Myo Plus has custom programs
that enable the clinician to select from a wide range of
Pulses widths {µS} Frequencies {Hz}, work/ rest, ramp up
and ramp down, and overall times. Also up to 5 phases per
program can be set up, for example: 5, 10, 15, 20, 30, 40
Hz, each frequency with its own pulse width and ramp up/down
and phase times.
|
|
| |
|
|
|
|
| |
| |
Copyright© 2005 Verity Medical Ltd. All Rights reserved
Revised: July 03, 2007
|
ISO9001:2000  ISO13485:2003
All our products are manufactured in compliance with the Medical Device Directive MDD93/42/EEC under the supervision of SGS, Notified Body number 0120. |
|
|