Verity Medical Logo

Detailed specifications

 

     
     
Products  

NeuroTrac™ ETS

Click here to return to the home pageClick here to view our TENS, Stimulators, Probes and EMG unitsClick here to view our range of Skin Electrodes and information on types and sizes of Skin ElectrodesClick here to view EMG, TENS, STIM and product FAQ'sClick here to view detailed information on TENS, Stimulation (STIM or EMS) and EMGClick here to view a list of Protocols developed by trained clinicians around the worldClick here to view retail and distributor sales, OEM Branding and custom product designClick here to download photos, brochures, manuals and protocols

OEM TENS, STIM and EMG products

 

 

NeuroTrac ETS

 

NeuroTrac ETS rear view

 

 

 

 

Specifications

EMG
1.0 Dual Channel
1.1 EMG Range: 0.2 to 2000 µV RMS (continuous)
1.2 Sensitivity: 0.l µV RMS
1.3 Accuracy: 4% of µV reading +/-0.3 µV at 200 Hz
1.4 Selectable Band pass filter - 3db Bandwidth,
        a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading below 235 microvolts
            10 Hz +/-3 Hz to 370 Hz +/- 10% - Reading above 235 microvolts
        b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
1.5 Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
1.6 Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
1.7 Battery: PP3 Alkaline
1.8 Work / Rest periods: 2-99 seconds
1.9 Number of Trials: 1-99

Neuromuscular Stimulation
2.1 Dual Channel: Individually isolated circuits
2.2 Amplitude: 0-80 mA: Indication only: actual mA will tend to be less than indicated due to Electrode impedance.
2.3 Type: Constant current
2.4 Waveform: Asymmetrical, rectangular, bi-phasic with net zero DC current
2.5 Pulse width selection: 50-450 µS (2% accuracy)
2.6 Pulse rate selection: 2-100 Hz (2% accuracy)
2.7 Output: selectable Synchronous or Alternating between channels A & B
2.8 Work / Rest periods: 2-60 seconds
2.9 Time 1 - 180 minutes
2.10 Ramp up time: 0.1 - 9.9 seconds
2.11 Preset and user programmable treatment Programs
2.12 Automatic output shut off with detection of open electrode above 11 mA

Low battery indication at 7.4 volts +/- 0.2 volts and automatic shut off 4 minutes after last key pressed, unless infra-red is turned on.

Environmental conditions for storage & transport: -10 to +50 degrees Centigrade.
0-90% Humidity.

Physical
Dimensions: Length 134mm, Width 69mm, Depth 29.7 mm

Weight
0.18 Kg